Error Messages
Perform operating panel diagnostics
Power supply voltage
for operating panel too low
LED -16-LD1 not lit Check fuse -22-F47
Ch
eck fuse -22-F48
Power supply voltage
central electrical system defective
LED -22-LD62 not lit Check fuse -22-F62
LED -22-LD64 not lit Check fuse -22-F64
Wiring defective Check wiring
Battery dead Charge battery, check battery acid,
replace battery
Operating panel
Message Meaning P
ossible cause Check Remedy
Error: 12 Volt
power supply voltage
too low
124 CU undervoltage
12V
Charge indicator lamp defective Check charge indicator lamp,
Check wiring
Dynamo defective Replace the dynamo
Internal error, operating panel Replace operating panel
Perform operating panel diagnostics
The controller of the dynamo is
defective
Replace the dynamo
Dynamo defective Replace the dynamo
Battery change-over relay defective Replace battery change-over relay
Internal error, operating panel Replace operating panel
CU overvoltage
12V
Error: 12 Volt
power supply voltage
too high
125
Zobrazit stránku 423