Bio-rad ChromLab™ Software Uživatelský manuál Strana 293

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 300
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 292
Performing Multicolumn Purification Chromatography
User Guide | 291
Performing Multicolumn Purification Chromatography
To perform multicolumn purification chromatography on an NGC system you must
Plumb the NGC system
Prime the NGC system
Select and save a Method template
Run the method
Plumbing the NGC System
How you set up and plumb the NGC system depends on the type of method you
plan to run. This section provides information about plumbing the system to use the
multicolumn purification templates.
Plumbing NGC Discover Pro Systems
This table lists the tasks for plumbing NGC Discover™ Pro systems. See Chapter 5,
Plumbing and Priming the NGC Chromatography Systems, in the NGC
Chromatography Software Installation Guide for detailed instructions.
Plumbing NGC Discover Pro Systems
Task
1.
Tube #6 between the outlet port on the sample inlet valve and the
inlet port on the sample pump
2.
Connect one of the following:
Tube #4 from the Common inlet port on the BioFrac fraction
collector diverter valve to port 1 on the outlet valve
Waste tubing to port 1 on the outlet valve
3.
Inlet tubing to the remaining ports on the sample inlet valve
4.
Outlet tubing to the remaining ports on the outlet valve
Note: Remove the inlet tubing from the sample pump inlet
manifold if it is installed.
Zobrazit stránku 292

Komentáře k této Příručce

Žádné komentáře