Running an Experiment
User Guide | 107
7. Double-click the System Pump module and choose the following settings in the
dialog box that appears:
a. If a buffer blending module is in line, select a Mode (Isocratic or Gradient)
under Flow Control.
b. Under Pressure Limits, select the lower and upper pressure limits.
c. (Optional) Select Control Flow to avoid overpressure. This setting causes
the flow rate to decr
ease automatically if the system exceeds set pressure
limits.
d. (Optional) Select Enable Air Sensor if air sensors are in line and set to
detect end of buf
fer.
Important: Select flow control settings for the run before you click Start.
Clicking Start starts the run.