6 | Creating a Method
184 | NGC Chromatography Systems and ChromLab Software
checkbox is also cleared. The other steps remain selected and will be
included in the scout.
Page 3: Generate Scout Sequence
On this page you set the number of runs to be performed, name each run, and set
the parameters for the scouted steps.
Number of runssets the number of runs to be performed as part of the
scout experiment.
Starting Valuesets the starting value for the parameter chosen to be
scouted. The default is the starting value in the base method.
Increment Value — sets the scout run increment values. The scout
increment value can be positive or negative.
The scout sequence table comprises the following columns:
Run # — the scout run number. This field is not editable.
Zobrazit stránku 185