Detectors
Instrument Guide | 6 9
Monitors up to four wavelengths in the UV/Vis range of 190–800 nm
Tip: Use ChromLab to change the absorbance wavelength values. For
more information, see the NGC Chromatography Systems and ChromLab
Software User Guide.
UV absorbance range: 0–3,000 mAU
UV linearity: ±5%
Three interchangeable flow cells with path lengths of 2 mm, 5 mm (default),
and 10 mm; see Table 4 for specific information
For information about changing the flow cells, see Replacing the UV Flow Cell
on page 196.
Note: Use the preparative 2 mm flow cell to measure high concentrations
of protein. The path lengths of the analytical 5 mm and 10 mm flow cells
are longer and are appropriate for measuring lower concentrations of
protein.
Conductivity Detector
Detailed Attributes
Factory specified range: 0.01–999.9 mS/cm
Validated working range: 0.01–300 mS/cm
Accuracy: ±2%
Flow cell volume: 6 μl
Integrated temperature sensor range: 4–100°C
Table 4. Pressure limits of UV detector flow cells
Pump Size Pressure 2 mm 5 mm 10 mm
F10 500 psi

F100 500 psi
