8
3. In the Calibration protocol name field, enter
OQ Mass Cal.
4. Select [Arg8]-Vasopressin (1084.247) from the drop-down list, and click Add
(keep the default values of 3% of expected peaks and minimum signal/noise
of 5).
5. Select Dynorphin A (2147.5) from the drop-down list, and click Add (keep the
default value of 3% of expected peaks and signal/noise of 5).
6. Select Beta-endorphin (3465.0) from the drop-down list, and click Add (keep
the default value at 3% of expected peaks and signal/noise of 5).
7. Select Arg-insulin (5963.8) from the drop-down list, and click Add (keep the
default value at 3% of expected peaks and signal/noise of 5).
8. Select Fit all 3 coefficients. Click OK.
10010680A.qxp 8/17/2007 1:58 PM Page 8