ProteinChip
®
OQ Kit
Instruction Manual and
Documentation
Catalog #C70-00080
For technical support call your local Bio-Rad office, or in the US, call 1-800-4BIORAD (1-800-424-6723)
For use with the ProteinChip SELDI system, Personal or Enterprise Edition,
with embedded system processor (ESP) version 1.1.15 or higher
10010680A.qxp 8/17/2007 1:58 PM Page FC1
Zobrazit stránku 0
1 2 3 4 5 6 ... 34 35

Shrnutí obsahu

Strany 1 - Catalog #C70-00080

ProteinChip®OQ KitInstruction Manual andDocumentationCatalog #C70-00080For technical support call your local Bio-Rad office, or in the US, call 1-800-

Strany 2

Step 4Create a StandardMass CalibrationRoutine76. In the Explorer pane of ProteinChip data manager software, select Projectsand click on any spectrum

Strany 3 - Table of Contents

83. In the Calibration protocol name field, enter OQ Mass Cal.4. Select [Arg8]-Vasopressin (1084.247) from the drop-down list, and click Add(keep the

Strany 4 - Introduction

2.3 OQ Maintenance Procedures 2.3.1 Maintenance Procedure 1 –– High Voltage (HV) Conditioning This procedure helps to decontaminate the surfaces of t

Strany 5

1. Insert the ProteinChip detector calibration array into the instrument.Note: For the Enterprise Edition instrument, place the array into slot 1 of a

Strany 6 - 1.2 Storage and Handling

7. The time required to complete this procedure is variable, and it may require up to severalhours as the instrument continually collects data of a sp

Strany 7 - Use of the ProteinChip OQ Kit

2.4 OQ Tests2.4.1 Test 1 –– Detector Sensitivity This test uses the ProteinChip detector qualification array to measure the signal-to-noise ratio(S/

Strany 8 - 2.2 File and Protocol Setup

6. Plot the peaks with noise set to 12,000 Da:a. Select all 8 spectra.b. Click on the data analysis icon . The Analysis Settings dialog opens.c. Open

Strany 9 - Conditioning Routine

7. Plot all 8 spectra and select the IgG peak using the peak selection tool.Note: Spots with high concentrations (spots A, C, E, and G) alternate with

Strany 10 - Mass Calibration

2.4.2 Test 2 — Mass Drift and Resolution 5.96 kD This test uses the ProteinChip peptide standard array to measure the mass drift and resolutionof ins

Strany 11 - OQ Mass Cal

6. Run the same protocol on partition “n” of 4 for all 8 spots. Record the partition number on the usage chart provided with the product packaging.7.

Strany 12

NoticesNo part of this publication may be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopy, record

Strany 13

8. Plot the peaks with the filtering option turned off:a. Select all 8 spectra. (Do not select the first laser optimization spectrum.)b. Click the ana

Strany 14

9. Plot all 8 spectra. Using the peak selection tool, select the insulin peak (5.963 kD) in all spectra.10. Export the data following the instructions

Strany 15 - 2.4 OQ Tests

2.4.3 Test 3 –– Resolution at 1 kDThis test uses the ProteinChip peptide standard array to measure the resolution of Arg-8-vasopressin.To do this, a

Strany 16

8. Plot the peaks with the filtering option turned off:a. Select all 8 spectra.b. Click the data analysis icon . The Analysis Settings dialog opens.c

Strany 17

9. Plot all 8 spectra (do not select the first optimization spectrum). Using the peak selectiontool, select the Arg-8-vasopressin (1,084.247 Da) peak.

Strany 18

2.4.4 Test 4 — Mass Accuracy This test uses the ProteinChip peptide standard array to test the mass accuracy of the system’sinternal and external cal

Strany 19

7. Plot the peaks with the filtering option turned off:a. Select all 8 spectra.b. Click the data analysis icon . The Analysis Settings dialog opens.c

Strany 20

Note: To ensure correct placement of the peak marker, zoom in on each peak during selection.Though some plots appear acceptable when zoomed out, zoomi

Strany 21

2510010680A.qxp 8/17/2007 1:59 PM Page 25

Strany 22

9. Zoom in and check that all 7 peaks in all 8 spectra are marked correctly. Failure to do somay result in incorrect results. Close the plot.Mass Accu

Strany 23

Table of ContentsChapter 1 Introduction ...11.1 Product and P

Strany 24

9. In the External Table, right-click on cell B3 and select Refresh Data.10. The spreadsheet indicates if the test passed or failed. The test passes i

Strany 25

12. Click OK. The spectrum flag changes from yellow to green for all 8 spectra, once calibrationis complete.13. Export the data following the instruct

Strany 26

Appendix A2910010680A.qxp 8/17/2007 1:59 PM Page 29

Strany 27

Appendix BProtocol Acceptance FormBio-Rad Laboratories recommends that the operational qualification (OQ) protocols be performedin total on a regular

Strany 28

Appendix CData Export ProcedureUse this procedure to export the data obtained during testing to the Microsoft Excel spreadsheetprovided with the kit (

Strany 29 - Spectrum

Life ScienceGroup00-0000 0000 Sig 1106Bulletin 0000 US/EG Rev ABio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 4BIORAD Austra

Strany 30

Chapter 1IntroductionQualification of analytical instruments is a formal process of documenting that an instrument isfit for its intended use and that

Strany 31

Completion of each maintenance procedure and test is tracked using the ProteinChip SELDI OQform, a Microsoft Excel spreadsheet supplied on the CD. Tes

Strany 32 - Appendix A

Table 1. Summary of procedures and tests (continued).RReeccoommmmeennddeedd CCoonnssuummaabbllee EEssttiimmaatteedd TTiimmee TTeesstt FFrreeqquuee

Strany 33 - Protocol Acceptance Form

Chapter 2Use of the ProteinChip OQ Kit2.1 OverviewThis chapter outlines the protocols used to perform the procedures and tests that comprise theProte

Strany 34 - Data Export Procedure

2.2 File and Protocol SetupComplete this procedure the first time you run the ProteinChip OQ kit. This procedure sets upthe file structure necessary

Strany 35 - Laboratories, Inc

Step 3Set Up theHigh Voltage (HV)Conditioning Routine61. Insert a blank ProteinChip array into the reader. If blank arrays are notavailable, use expir