NGC™ Chromatography Systems and ChromLab™ SoftwareInstallation GuideVersion 3.03.0

Strany 2

1 | Introduction8 | NGC Chromatography Systems and ChromLab SoftwareMain FeaturesNGC chromatography systems enable you to do the following:

Strany 3 - ChromLab™ Software

5 | Plumbing and Priming the NGC Chromatography Systems98 | NGC Chromatography Systems and ChromLab SoftwarePreparing the Buffer and Waste

Strany 4

Plumbing NGC SystemsInstallation Guide | 9 9Plumbing NGC SystemsSeveral plumbing tasks are required for all instrument configurations. Other tasks

Strany 5 - Table of Contents

5 | Plumbing and Priming the NGC Chromatography Systems100 | NGC Chromatography Systems and ChromLab SoftwarePlumbing All NGC SystemsTable

Strany 6

Plumbing NGC SystemsInstallation Guide | 101System Pumps Wash LinesSample Pump Wash Line

Strany 7 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems102 | NGC Chromatography Systems and ChromLab SoftwareTo plumb and prime the system

Strany 8

Plumbing NGC SystemsInstallation Guide | 1038. Attach a 30 ml syringe to the 1/4-28 female to male luer adaptor.9. Withdraw the plunger on the syr

Strany 9 - 1 Introduction

5 | Plumbing and Priming the NGC Chromatography Systems104 | NGC Chromatography Systems and ChromLab SoftwarePlumbing the Sample Inject Va

Strany 10 - Main Features

Plumbing NGC SystemsInstallation Guide | 105Attaching the Backpressure RegulatorIf your column requires a high backpressure, install the backpress

Strany 11 - Safety Information

5 | Plumbing and Priming the NGC Chromatography Systems106 | NGC Chromatography Systems and ChromLab SoftwarePlumbing the NGC Quest System

Strany 12 - Regulatory Information

Plumbing NGC SystemsInstallation Guide | 107NGC Quest Point-to-Plumb Connections

Strany 13 - FCC Warning and Notes

Safety InformationInstallation Guide | 9Safety InformationThis section explains the different safety alerts within this document. This section als

Strany 14 - Finding Out More

5 | Plumbing and Priming the NGC Chromatography Systems108 | NGC Chromatography Systems and ChromLab SoftwareTo plumb the system pumps1. C

Strany 15 - 2 Preparing Your Site

Plumbing NGC SystemsInstallation Guide | 109Plumbing the NGC Scout SystemTable 10 lists the tasks for plumbing the NGC Scout™ system. The tasks a

Strany 16 - Item Specification

5 | Plumbing and Priming the NGC Chromatography Systems110 | NGC Chromatography Systems and ChromLab SoftwareNGC Scout Point-to-Plumb Conn

Strany 17 - Room Requirements

Plumbing NGC SystemsInstallation Guide | 111To plumb the buffer blending valve Connect the inlet tubing to ports Q1–Q4 on the buffer blending val

Strany 18 - Environment

5 | Plumbing and Priming the NGC Chromatography Systems112 | NGC Chromatography Systems and ChromLab Software3. Immerse the open end of th

Strany 19 - Power Requirements

Plumbing NGC SystemsInstallation Guide | 113Plumbing the NGC Discover SystemTable 11 lists the tasks for plumbing the NGC Discover™ system. The t

Strany 20 - ChromLab System Requirements

5 | Plumbing and Priming the NGC Chromatography Systems114 | NGC Chromatography Systems and ChromLab SoftwareNGC Discover Point-to-Plumb C

Strany 21 - Installation Guide

Plumbing NGC SystemsInstallation Guide | 115To plumb the buffer blending valve Connect the inlet tubing to ports Q1–Q4 on the buffer blending val

Strany 22 - To change the font size

5 | Plumbing and Priming the NGC Chromatography Systems116 | NGC Chromatography Systems and ChromLab SoftwareTo plumb the buffer inlet val

Strany 23 - Chromatography Systems

Plumbing NGC SystemsInstallation Guide | 117To plumb the pH valve1. Connect the pump head wash priming tubing to the Cal In port on the pH valve.2

Strany 24 - NGC Systems Accessories Kit

1 | Introduction10 | NGC Chromatography Systems and ChromLab SoftwareRegulatory InformationThe NGC instrument has been tested and found to b

Strany 25 - NGC Discover Accessories Kit

5 | Plumbing and Priming the NGC Chromatography Systems118 | NGC Chromatography Systems and ChromLab SoftwarePlumbing the NGC Discover Pro

Strany 26 - Attaching the Touch Screen

Plumbing NGC SystemsInstallation Guide | 119NGC Discover Pro Point-to-Plumb Connections

Strany 27 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems120 | NGC Chromatography Systems and ChromLab SoftwarePlumbing Sample Inlet ValvesE

Strany 28

Plumbing NGC SystemsInstallation Guide | 1212. If your system has two sample inlet valves:a. Connect tubing from the Out port on sample inlet valv

Strany 29 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems122 | NGC Chromatography Systems and ChromLab Softwareb. (Optional) Do one of the f

Strany 30

Priming and Purging the SystemInstallation Guide | 123Priming and Purging the SystemThis section explains how to prime your system before running

Strany 31 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems124 | NGC Chromatography Systems and ChromLab SoftwarePriming and Purging the NGC Q

Strany 32

Priming and Purging the SystemInstallation Guide | 1256. Remove the syringe from the priming port and discard the liquid into waste.7. Repeat step

Strany 33 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems126 | NGC Chromatography Systems and ChromLab Software4. In the fluidic scheme, tou

Strany 34

Priming and Purging the SystemInstallation Guide | 127Priming and Purging the NGC Scout SystemPriming the NGC Scout system is a manual procedure a

Strany 35 - Installation Guide

Regulatory InformationInstallation Guide | 1 1FCC Warning and NotesThe NGC instrument complies with part 15 of the FCC Rules. Operation is subject

Strany 36

5 | Plumbing and Priming the NGC Chromatography Systems128 | NGC Chromatography Systems and ChromLab Software5. Select Q4 as the port to o

Strany 37 - Installation Guide

Priming and Purging the SystemInstallation Guide | 129Purging the NGC Scout SystemTo purge the buffer blending valve 1. Open the System Pump dialo

Strany 38 - Stand kit (#788-5040)

5 | Plumbing and Priming the NGC Chromatography Systems130 | NGC Chromatography Systems and ChromLab Software3. To purge all lines of air

Strany 39 - Installation Guide

Priming and Purging the SystemInstallation Guide | 131Priming and Purging the NGC Discover SystemPriming the NGC Discover system is a manual proce

Strany 40

5 | Plumbing and Priming the NGC Chromatography Systems132 | NGC Chromatography Systems and ChromLab Software4. In the fluidic scheme, tou

Strany 41 - Installation Guide

Priming and Purging the SystemInstallation Guide | 1337. Insert a 30 ml syringe into the priming port on system pump A.8. Turn the priming port va

Strany 42

5 | Plumbing and Priming the NGC Chromatography Systems134 | NGC Chromatography Systems and ChromLab SoftwareTo prime the buffer inlet val

Strany 43 - Installation Guide

Priming and Purging the SystemInstallation Guide | 1359. Repeat steps 4–8 for the remaining buffer inlet ports.10. To prime buffer inlet valve B,

Strany 44 - To install the pH probe

5 | Plumbing and Priming the NGC Chromatography Systems136 | NGC Chromatography Systems and ChromLab SoftwareTo prime the sample inlet val

Strany 45 - Installation Guide

Priming and Purging the SystemInstallation Guide | 1375. Insert a 30 ml syringe into the sample pump’s priming port.6. Slowly withdraw the plunger

Strany 46

1 | Introduction12 | NGC Chromatography Systems and ChromLab SoftwareFinding Out MoreAfter you install NGC documentation from the NGC Chroma

Strany 47 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems138 | NGC Chromatography Systems and ChromLab SoftwarePurging the NGC Discover Syst

Strany 48 - Previously attached

Priming and Purging the SystemInstallation Guide | 1392. To purge an individual line of air bubbles and fill it with buffer:a. Select the port to

Strany 49 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems140 | NGC Chromatography Systems and ChromLab SoftwareTo purge the buffer inlet val

Strany 50

Priming and Purging the SystemInstallation Guide | 141To purge the sample pump1. In the fluidic scheme, touch or click the Sample Pump module to o

Strany 51 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems142 | NGC Chromatography Systems and ChromLab Software3. Click Change Position to s

Strany 52 - Task: Refer to the section:

Priming and Purging the SystemInstallation Guide | 1433. Set Flow Rate to 10 ml/min.4. Click Change Position to set the sample inject valve to Sam

Strany 53 - Inserting Additional Modules

5 | Plumbing and Priming the NGC Chromatography Systems144 | NGC Chromatography Systems and ChromLab SoftwareCalibrating the Instrument Th

Strany 54

Calibrating the InstrumentInstallation Guide | 145Calibrating the Pump Flow RatePrime and purge the system pumps with water before starting the ca

Strany 55 - Installation Guide

5 | Plumbing and Priming the NGC Chromatography Systems146 | NGC Chromatography Systems and ChromLab SoftwareCalibrating the Conductivity

Strany 56

Installation Guide | 147A Upgrading ChromLab Software and NGC SystemsThis appendix explains how to upgrade ChromLab™ software and your NGC™ system

Strany 57 - Installation Guide

Installation Guide | 1 32 Preparing Your SiteThis chapter explains how to prepare your site to successfully install your NGC™ chromatography syste

Strany 58 -  Connecting the Air Sensors

A | Upgrading ChromLab Software and NGC Systems148 | NGC Chromatography Systems and ChromLab SoftwareUpdating NGC SystemsUpdating the NGC

Strany 59 - Installation Guide

Updating NGC SystemsInstallation Guide | 1493. After the update process completes, the instrument automatically reboots. The system checks the fir

Strany 60

A | Upgrading ChromLab Software and NGC Systems150 | NGC Chromatography Systems and ChromLab Software

Strany 61 - Installation Guide

Installation Guide | 151B Troubleshooting ChromLab Software InstallationThis appendix explains what to do if ChromLab™ software fails to connect t

Strany 62 -  UV monitor

B | Troubleshooting ChromLab Software Installation152 | NGC Chromatography Systems and ChromLab SoftwareTable 14. Possible Causes and Sol

Strany 63 - Installation Guide

Possible Causes for Connection FailuresInstallation Guide | 153A firewall on the ChromLab computer is turned on and does not allow the ChromLab pr

Strany 64 - Connecting the Air Sensors

B | Troubleshooting ChromLab Software Installation154 | NGC Chromatography Systems and ChromLab SoftwareConfiguring the Network Adapter fo

Strany 65 - (air sensor to pump inlet)

Configuring the Network Adapter for Direct ConnectionInstallation Guide | 1553. Review the list of network connections. If you see at least one ic

Strany 66

B | Troubleshooting ChromLab Software Installation156 | NGC Chromatography Systems and ChromLab Software5. Right-click the Local Area Conn

Strany 67 - To activate the air sensors

Configuring the Network Adapter for Direct ConnectionInstallation Guide | 157Note: If you do not see this item listed, the computer is not set up

Strany 68

2 | Preparing Your Site14 | NGC Chromatography Systems and ChromLab SoftwareNGC Scout Plus Length: 32 in.; 81.3 cmWidth: 27 in.; 68.6 cmHei

Strany 69 - Installation Guide

B | Troubleshooting ChromLab Software Installation158 | NGC Chromatography Systems and ChromLab SoftwareTroubleshooting Network Connection

Strany 70

Troubleshooting Network Connection IssuesInstallation Guide | 159IP Address Located in the System Information Dialog BoxIf the system is running,

Strany 71 - Installation Guide

B | Troubleshooting ChromLab Software Installation160 | NGC Chromatography Systems and ChromLab SoftwareChecking the Network ConnectionAft

Strany 72

Troubleshooting Network Connection IssuesInstallation Guide | 161 If the network is not set up correctly, you see results similar to this image:T

Strany 73 - Installation Guide

B | Troubleshooting ChromLab Software Installation162 | NGC Chromatography Systems and ChromLab SoftwareChromLab Software Firewall Setting

Strany 74

Firmware Update FailureInstallation Guide | 163Firmware Update FailureIf the firmware upgrade fails, restart the NGC instrument. The software atte

Strany 75 - Installation Guide

B | Troubleshooting ChromLab Software Installation164 | NGC Chromatography Systems and ChromLab Software

Strany 76

Installation Guide | 165C Connecting the C-96 Autosampler to NGC SystemsUsing the C-96 Autosampler with NGC SystemsThe C-96 autosampler enhances N

Strany 77 - Positive

C | Connecting the C-96 Autosampler to NGC Systems166 | NGC Chromatography Systems and ChromLab Software5. Locate the two cables that conn

Strany 78

Using the C-96 Autosampler with NGC SystemsInstallation Guide | 1678. On the back of the autosampler:a. Insert the 15-pin cable into the left (P1

Strany 79 - Installation Guide

Room RequirementsInstallation Guide | 1 5Room RequirementsBench SpaceEnsure that the bench or rack on which you install the NGC system is designed

Strany 80

C | Connecting the C-96 Autosampler to NGC Systems168 | NGC Chromatography Systems and ChromLab SoftwarePlumbing and Priming the C-96 Auto

Strany 81 - Installation Guide

Using the C-96 Autosampler with NGC SystemsInstallation Guide | 169To prime and purge the C-96 autosamplerNote: Before priming the autosampler, e

Strany 82

C | Connecting the C-96 Autosampler to NGC Systems170 | NGC Chromatography Systems and ChromLab Software6. Set the inject valve to System

Strany 83 - Software

Using the C-96 Autosampler with NGC SystemsInstallation Guide | 171

Strany 84 - Installing ChromLab Software

C | Connecting the C-96 Autosampler to NGC Systems172 | NGC Chromatography Systems and ChromLab SoftwareSetting Up the C-96 Autosampler fo

Strany 85 - The wizard installs ChromLab

Installation Guide | 173D Firewall ConfigurationFirewall Inbound Rules for Bio-Rad NGC SoftwareBio-Rad configures the ChromLab™ software using Mic

Strany 86 - Software and the NGC Systems

D | Firewall Configuration174 | NGC Chromatography Systems and ChromLab SoftwareFor all services and applications listed in Table 15, the f

Strany 88 - Setting the System Name

Life ScienceGroup14-0273 0114 Sig 121310026252 Rev C US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61

Strany 89 - Installation Guide

2 | Preparing Your Site16 | NGC Chromatography Systems and ChromLab SoftwareEnvironment Operating conditions 4 to 40°C(35 to 104°F)0–95% RH,

Strany 90 - On the ChromLab Computer

Power RequirementsInstallation Guide | 1 7Power RequirementsThe power supply to the NGC system must be stable and within specifications at all tim

Strany 92 - On the NGC Instrument

2 | Preparing Your Site18 | NGC Chromatography Systems and ChromLab SoftwareChromLab System RequirementsBio-Rad can provide a PC with ChromL

Strany 93 - Installation Guide

Preparing the ChromLab ComputerInstallation Guide | 1 9Preparing the ChromLab ComputerTo ensure the ChromLab computer operates the NGC system smoo

Strany 94

2 | Preparing Your Site20 | NGC Chromatography Systems and ChromLab SoftwareTo turn off automatic updates1. On the ChromLab computer, select

Strany 95 - Setting Delay Volume

Installation Guide | 2 13 Setting Up NGC Chromatography SystemsAfter you unpack the NGC™ chromatography system and install it in a standard cold c

Strany 96

3 | Setting Up NGC Chromatography Systems22 | NGC Chromatography Systems and ChromLab SoftwareNGC Systems Accessories KitEach NGC system s

Strany 97 - Installation Guide

NGC Systems Accessories KitInstallation Guide | 2 3NGC Scout Accessories KitThe kit for the NGC Scout™ system includes all the accessories in the

Strany 98

3 | Setting Up NGC Chromatography Systems24 | NGC Chromatography Systems and ChromLab SoftwareAttaching the Touch ScreenThe installation p

Strany 99 - 5 Plumbing and Priming the

Attaching the Touch ScreenInstallation Guide | 2 5Note: The top section in this image is a third tier. (The NGC Discover system is a 3-tier syste

Strany 100 - Waste Bottles

3 | Setting Up NGC Chromatography Systems26 | NGC Chromatography Systems and ChromLab Software4. Using two ¼-20 x ¾" hex bolts and th

Strany 101 - Plumbing NGC Systems

Attaching the Touch ScreenInstallation Guide | 2 77. Locate the touch screen display cable in the accessories kit.8. Firmly push the touch screen

Strany 102 - Plumbing All NGC Systems

NGC™ Chromatography Systems and ChromLab™ SoftwareInstallation GuideVersion 3.0

Strany 103 - System Pumps Wash Lines

3 | Setting Up NGC Chromatography Systems28 | NGC Chromatography Systems and ChromLab Software9. Route the display cable through the side

Strany 104

Attaching the Touch ScreenInstallation Guide | 2 911. Locate the touch screen cable cover in the accessories kit. This is the cover with the cable

Strany 105 - To plumb the sample pump

3 | Setting Up NGC Chromatography Systems30 | NGC Chromatography Systems and ChromLab Software13. Secure the cover to the back of the inst

Strany 106

Attaching the Touch ScreenInstallation Guide | 3 1Repositioning the Touch Screen Pivot Arm AssemblyIf you attach the touch screen to the right sid

Strany 107 - Installation Guide

3 | Setting Up NGC Chromatography Systems32 | NGC Chromatography Systems and ChromLab Software3. Using a Phillips screwdriver, remove the

Strany 108 - Plumbing the NGC Quest System

Attaching the Touch ScreenInstallation Guide | 3 39. After you attach the mounting post to the side of the instrument, slide the touch screen disp

Strany 109 - Installation Guide

3 | Setting Up NGC Chromatography Systems34 | NGC Chromatography Systems and ChromLab SoftwareNGC System Location Is a Standard Cold Cabin

Strany 110 - To plumb the system pumps

Attaching the Touch ScreenInstallation Guide | 3 54. Complete the steps in the section To connect the power and communication cables on page 77.T

Strany 111 - Plumbing the NGC Scout System

3 | Setting Up NGC Chromatography Systems36 | NGC Chromatography Systems and ChromLab Software3. Using a Phillips screwdriver, remove the

Strany 112

Attaching the Touch ScreenInstallation Guide | 3 75. Position the bracket on the back of the touch screen so that the bracket opening is on the bo

Strany 113 - To plumb the pH valve

Bio-Rad Technical Support DepartmentThe Bio-Rad Technical Support department in the U.S. is open Monday through Friday, 5:00 AM to 5:00 PM, Pacific Ti

Strany 114

3 | Setting Up NGC Chromatography Systems38 | NGC Chromatography Systems and ChromLab Software7. Locate the four thumb screws in the packa

Strany 115 - Installation Guide

Connecting the Drain Tube to the Buffer TrayInstallation Guide | 3 9Connecting the Drain Tube to the Buffer TrayThe top of the NGC system is molde

Strany 116

3 | Setting Up NGC Chromatography Systems40 | NGC Chromatography Systems and ChromLab Software4. Uncoil the drain tube and insert it throu

Strany 117 - Tubing color Port Solution

Connecting the Drain Tube to the Buffer TrayInstallation Guide | 4 16. Turn the top right side up and place it on the NGC instrument.7. Place the

Strany 118

3 | Setting Up NGC Chromatography Systems42 | NGC Chromatography Systems and ChromLab SoftwareInstalling the pH Probe in the pH ModuleNote

Strany 119

Installing the pH Probe in the pH ModuleInstallation Guide | 4 310. Start the NGC instrument.The following figure shows the pH probe installed in

Strany 120

3 | Setting Up NGC Chromatography Systems44 | NGC Chromatography Systems and ChromLab SoftwareAttaching an Expansion Tier to the NGC Instr

Strany 121 - Installation Guide

Attaching an Expansion Tier to the NGC InstrumentInstallation Guide | 4 5Tip: The buffer tray can remain inside the top.4. Locate the package th

Strany 122 - Plumbing Sample Inlet Valves

3 | Setting Up NGC Chromatography Systems46 | NGC Chromatography Systems and ChromLab Software6. Place the expansion tier on top of the al

Strany 123 - Plumbing Outlet Valves

Attaching an Expansion Tier to the NGC InstrumentInstallation Guide | 4 78. View the connector ends. Each end is beveled. The direction of the bev

Strany 124

Installation Guide | iiiTable of ContentsChapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7Ma

Strany 125 - Installation Guide

3 | Setting Up NGC Chromatography Systems48 | NGC Chromatography Systems and ChromLab Software9. Firmly press the clips inward and insert

Strany 126 - To prime the system pumps

Attaching an Expansion Tier to the NGC InstrumentInstallation Guide | 4 910. Insert the other connector into the lower tier’s connector port.11. P

Strany 127 - To purge the system pumps

3 | Setting Up NGC Chromatography Systems50 | NGC Chromatography Systems and ChromLab Software13. Perform the following tasks:Task: Refer

Strany 128

Inserting Additional ModulesInstallation Guide | 5 1Inserting Additional ModulesWARNING! Disconnect power to the NGC instrument before inserting

Strany 129 - Priming the NGC Scout System

3 | Setting Up NGC Chromatography Systems52 | NGC Chromatography Systems and ChromLab Software5. (Optional) Convert the bay to a single-wi

Strany 130

Inserting Additional ModulesInstallation Guide | 5 310. Insert the screws that shipped with the module into the screw holes.Caution: Do not overti

Strany 131 - Purging the NGC Scout System

3 | Setting Up NGC Chromatography Systems54 | NGC Chromatography Systems and ChromLab SoftwareTo convert a single-wide bay to a double-wid

Strany 132

Inserting Additional ModulesInstallation Guide | 5 5To convert a double-wide bay to a single-wide bay1. Complete steps 1–4 in the section To inser

Strany 133 - Priming NGC Discover Systems

3 | Setting Up NGC Chromatography Systems56 | NGC Chromatography Systems and ChromLab SoftwareConnecting Peripheral EquipmentThis section

Strany 134

Connecting Peripheral EquipmentInstallation Guide | 5 73. Insert system cable 17 that ships with the fraction collector into an INSTRUMENT BUS con

Strany 135 - Installation Guide

iv | NGC Chromatography Systems and ChromLab SoftwareTable of ContentsAttaching the Touch Screen . . . . . . . . . . . . . . . . . . . . . . . .

Strany 136

3 | Setting Up NGC Chromatography Systems58 | NGC Chromatography Systems and ChromLab Software6. Insert one end of the peripheral cable in

Strany 137 - To prime the sample pump

Connecting Peripheral EquipmentInstallation Guide | 5 98. Power on the instruments:a. Power on the BioFrac fraction collector first.b. Press the s

Strany 138

3 | Setting Up NGC Chromatography Systems60 | NGC Chromatography Systems and ChromLab SoftwareConnecting Tubing to the ColumnsNote: Ensur

Strany 139 - Installation Guide

Connecting Peripheral EquipmentInstallation Guide | 6 1Connecting a Keyboard or MouseThe NGC systems can be controlled using the keyboard on the t

Strany 140

3 | Setting Up NGC Chromatography Systems62 | NGC Chromatography Systems and ChromLab SoftwareConnecting the Air SensorsThe system support

Strany 141 - Installation Guide

Connecting Peripheral EquipmentInstallation Guide | 6 3Air Sensor Tubing Length and Bubble SizeThe response of the air sensor module is highly dep

Strany 142

3 | Setting Up NGC Chromatography Systems64 | NGC Chromatography Systems and ChromLab SoftwareAttaching and Activating Air SensorsThis sec

Strany 143 - To purge the sample pump

Connecting Peripheral EquipmentInstallation Guide | 6 55. To connect an air sensor to a system pump:a. Connect tubing from the buffer bottle or fr

Strany 144

3 | Setting Up NGC Chromatography Systems66 | NGC Chromatography Systems and ChromLab SoftwareConnecting External Devices to the NGC Instr

Strany 145 - Installation Guide

Connecting Peripheral EquipmentInstallation Guide | 6 7Importing Analog Signals to the NGC InstrumentYou can import analog (voltage) signal from u

Strany 146 - Calibrating the Instrument

Installation Guide | vTable of ContentsPlumbing NGC Systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 147 - To zero the pressure

3 | Setting Up NGC Chromatography Systems68 | NGC Chromatography Systems and ChromLab Software3. Insert the other end of the cable into on

Strany 148

Connecting Peripheral EquipmentInstallation Guide | 6 9Connecting External Devices to the SIMWARNING! These procedures require you to work with

Strany 149 - A Upgrading ChromLab

3 | Setting Up NGC Chromatography Systems70 | NGC Chromatography Systems and ChromLab Software3. Loosen all three screws on the top of the

Strany 150 - Updating NGC Systems

Connecting Peripheral EquipmentInstallation Guide | 7 19. On the analog output cable, determine the positive and negative wires.10. Insert the pos

Strany 151 - Installation Guide

3 | Setting Up NGC Chromatography Systems72 | NGC Chromatography Systems and ChromLab Software11. Insert the negative wire and one end of

Strany 152

Connecting Peripheral EquipmentInstallation Guide | 7 313. Insert the connector into the SIM 1 or SIM 2 port on the back of the SIM.14. Refer to t

Strany 153 - B Troubleshooting ChromLab

3 | Setting Up NGC Chromatography Systems74 | NGC Chromatography Systems and ChromLab SoftwareTo connect an external device to the DAC A a

Strany 154 - Possible Cause Solution

Connecting Peripheral EquipmentInstallation Guide | 7 59. Insert the positive and ground wires into the appropriate terminals on the DAC A connec

Strany 155

3 | Setting Up NGC Chromatography Systems76 | NGC Chromatography Systems and ChromLab SoftwareConnecting the Power and Communication Cable

Strany 156

Connecting the Power and Communication CablesInstallation Guide | 7 7To connect the power and communication cables1. If you have not already done

Strany 157 - Installation Guide

vi | NGC Chromatography Systems and ChromLab SoftwareTable of ContentsAppendix C Connecting the C-96 Autosampler to NGC Systems . . . . . . . .

Strany 158

3 | Setting Up NGC Chromatography Systems78 | NGC Chromatography Systems and ChromLab SoftwareConnecting the NGC System to the ChromLab Co

Strany 159 - Installation Guide

Connecting the NGC System to the ChromLab ComputerInstallation Guide | 7 9Establishing a Network ConnectionWhen establishing a connection between

Strany 160

3 | Setting Up NGC Chromatography Systems80 | NGC Chromatography Systems and ChromLab Software

Strany 161 - Installation Guide

Installation Guide | 8 14 Installing ChromLab SoftwareThis chapter explains how to install ChromLab™ software on your computer.If this is the firs

Strany 162

4 | Installing ChromLab Software82 | NGC Chromatography Systems and ChromLab SoftwareInstalling ChromLab SoftwareTo install ChromLab softwar

Strany 163 - Installation Guide

Installing ChromLab SoftwareInstallation Guide | 8 3The wizard installs ChromLab.Note: Installing ChromLab can take several minutes.5. When the i

Strany 164 - NGC ChromLab Application

4 | Installing ChromLab Software84 | NGC Chromatography Systems and ChromLab SoftwareEstablishing the Connection Between ChromLab Software a

Strany 165 - Firmware Update Failure

Establishing the Connection Between ChromLab Software and the NGC SystemsInstallation Guide | 8 54. Press the soft power button on the side of the

Strany 166

4 | Installing ChromLab Software86 | NGC Chromatography Systems and ChromLab SoftwareSetting the System NameThe tasks in this section are pe

Strany 167 - C Connecting the C-96

Establishing the Connection Between ChromLab Software and the NGC SystemsInstallation Guide | 8 72. Touch or click OK.The ChromLab connection info

Strany 168

Installation Guide | 71 IntroductionThe NGC™ chromatography systems are preparative systems designed to rapidly automate the purification of biomo

Strany 169 - Installation Guide

4 | Installing ChromLab Software88 | NGC Chromatography Systems and ChromLab SoftwareOn the ChromLab Computer1. Start ChromLab.The System Co

Strany 170 - To prepare the tubing

Establishing the Connection Between ChromLab Software and the NGC SystemsInstallation Guide | 8 93. On the NGC instrument you see that the system

Strany 171 - Sample Inject

4 | Installing ChromLab Software90 | NGC Chromatography Systems and ChromLab SoftwareThere are several possible causes for connection to fai

Strany 172

Establishing the Connection Between ChromLab Software and the NGC SystemsInstallation Guide | 9 1Click Restart System to synchronize the system ti

Strany 173 - Installation Guide

4 | Installing ChromLab Software92 | NGC Chromatography Systems and ChromLab Software

Strany 174

Setting Delay VolumeInstallation Guide | 9 3Setting Delay VolumeThe delay volume is the volume of fluid in the path between the UV or UV/Vis detec

Strany 175 - D Firewall Configuration

4 | Installing ChromLab Software94 | NGC Chromatography Systems and ChromLab Software3. Select Synchronize with detector.4. Refer to the fol

Strany 176

Setting Delay VolumeInstallation Guide | 9 5tubing length between the UV outlet port and outlet valve as well as the tubing length between the out

Strany 177

4 | Installing ChromLab Software96 | NGC Chromatography Systems and ChromLab SoftwareSetting Up Remote Access ConnectionYou can access the N

Strany 178 - Laboratories, Inc

Installation Guide | 9 75 Plumbing and Priming the NGC Chromatography SystemsWith the exception of the inlet tubing, waste lines, and column, the