Bio-rad TC20™ Automated Cell Counter Uživatelský manuál

Procházejte online nebo si stáhněte Uživatelský manuál pro Měřící nástroje Bio-rad TC20™ Automated Cell Counter. Bio-Rad TC20™ Automated Cell Counter User Manual Uživatelská příručka

  • Stažení
  • Přidat do mých příruček
  • Tisk

Shrnutí obsahu

Strany 1 - Automated Cell Counter

Catalog #145-0101TC20™ Automated Cell CounterInstruction Manual

Strany 2

LCD screenKeys for navigationUSB ports A and BSlide portPower switchFig. 1. TC20 cell counter.Fig. 2. Home screen.3 IntroductionUser Interface The use

Strany 3 - Cell Counter

Fig. 3. Options screen.4IntroductionOptions — for selecting additional operationsOptions Screen The Options screen (Figure 3) contains the following c

Strany 4 - Warranty

Fig. 4. Counting slides and trypan blue dye.Fig. 5. Accessories for the TC20 cell counter.5IntroductionTrypan Blue Dye The Trypan blue dye has been sp

Strany 5 - FCC Warning and Notes

Chapter 2 — Setting Up the TC20™ Automated Cell Counter2

Strany 6

Fig. 6. Setting the date and time.8Setting Up the TC20 Automated Cell CounterSystem ComponentsThe TC20 automated cell counter (catalog #145-0102) incl

Strany 7 - Table of Contents

95. Finally, when you have made all the modifications use the up or down arrow key to reach the Yes or No options for the Save date and time? setting.

Strany 8 - Chapter 1 — Introduction

4. Finally, when you have made all the modifications use the up or down key to reach Yes or No options for the Save changes? setting. 5. Press Ente

Strany 9

3Chapter 3 — Performing Cell Counts

Strany 10 - User Interface

Fig. 9. Loading the counting slide.12Performing Cell CountsPerforming Cell CountsLoading SlidesHandle the TC20™ counting slides using the edges and av

Strany 11 - Counting Slides

Fig. 10. Inserting the counting slide into the TC20 cell counter.Fig. 11. Cell count screen without trypan blue dye.Fig. 12. Cell count screen with tr

Strany 12 - System Verification Kit

Copyright ©2011 Bio-Rad Laboratories, Inc. Reproduction in any form, either print or electronic, is prohibited without written permission of Bio-Rad L

Strany 13 - Chapter 2 —

14Counting Samples with Multiple Cell PopulationsWhen counting samples with multiple cell populations that vary greatly in size (for example, fibroblas

Strany 14 - Setting the Date and Time

15If automatic name serialization is enabled, you can select a name consisting of maximum six characters, the last two fields are reserved for automati

Strany 15 - Setting up Gating

Fig. 15. Image of cells.16Performing Cell CountsExporting the Image of Cells from the Current CountThe image from the most recent count can be downloa

Strany 16

17Performing Cell CountsFig. 16. Histogram for total cell count (sample without trypan blue dye).Fig. 17. Histogram for live and dead cell count (sam

Strany 17 - Performing Cell Counts

Chapter 4 — Saved Count Results4

Strany 18

Fig. 18. Previous counts results.20Saved Count ResultsSaved Count ResultsViewing Previous CountsThe TC20™ cell counter stores results from 100 reads,

Strany 19 - TC20 Automated Cell Counter

21Saved Count Results2. Use the down key to select Export previous counts, and press Enter. If the USB flash drive is already inserted into instrum

Strany 20 - Sample Naming

22Deleting All Counts To delete all counts stored in the instrument:1. From the Home screen, use the down arrow key to select Options, and press Enter

Strany 21 - Current Count Options

Chapter 5 — Maintenance and Troubleshooting5

Strany 22

Fig. 19. Disabling user defined gates.Fig. 20. System verification results.24Maintenance and TroubleshootingMaintenance and TroubleshootingPerforming th

Strany 23

TC20™ AutomatedCell CounterInstruction Manual

Strany 24 - Saved Count Results

Fig. 19. Disabling user defined gates.Fig. 20. System verification results.25Maintenance and TroubleshootingCleaning the TC20 Automated Cell CounterThe

Strany 25

26Maintenance and TroubleshootingTo load a new roll of labels:1. Turn the printer on. 2. Open the cover.3. Place the new roll as directed, with the fir

Strany 26 - Automatic Data Export

27 Maintenance and TroubleshootingTroubleshootingError Message Possible Cause and Solution Invalid result data Data were improperly stored and can

Strany 27 - Deleting All Counts

28Ordering InformationCatalog # DescriptionTC10 Automated Cell Counter145 -0102 TC20 Automated Cell Counter, 120–240 V, includes instrument, power s

Strany 28 - Troubleshooting

Life ScienceGroup13-1888 1213 Sig 121310024423 Rev B US/EGBio-Rad Laboratories, Inc.Web site www.bio-rad.com USA 800 424 6723 Australia 61

Strany 29

iiBio-Rad Laboratories ResourcesTable 1 lists Bio-Rad resources and how to locate what you need. Table 1. Bio-Rad resources.Resource How to Con

Strany 30 - Thermal Printer Maintenance

iiiSafety ComplianceThis instrument has been tested and found to be in compliance with all applicable requirements of the following safety and electro

Strany 31 - Updating the Firmware

ivBio-Rad Laboratories is not responsible for any injury or damage caused by the use of this instrument for purposes other than those for which it is

Strany 32

vTable of ContentsBio-Rad Laboratories Resources ...iiWarranty ...

Strany 33 - Ordering Information

Chapter 1 — Introduction1

Strany 34 - Laboratories, Inc

2IntroductionIntroductionThe Bio-Rad TC20™ automated cell counter is a benchtop instrument that counts mammalian cells in one simple step, initiating

Komentáře k této Příručce